How to Use Auxiliary Functions

NOTE
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com

How to Use Auxiliary Functions

Automatic Realignment

Automatic Realignment realigns the laser cavity after Laser Protection. You
should use Automatic Realignment if you have already tried to reactivate
the laser and to reduce power, and this has been unsuccessful.
To get the specified performance, ensure that the warm-up time has
passed before starting realignment.
To realign the laser cavity:
1 Move to the Tunable Laser channel and press Menu.
2 Move to <Realign> and press [OK]. You will see the screen in Figure 81 .
3 Wait approximately 30 minutes. This time depends on how much the
wavelength of the instrument has drifted since the last Automatic Realignment
was performed.
Figure 81 Realign Screen
Tunable Lasers
153