Status Listings
Analyzer Errors
Analyzer Errors
Table 7-10. Analyzer Errors (1 of 2)
Error Number
20001
20002
20003
20005
20007
7-38
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Description [description/explanation/examples]
desc = "Error detected in ADC sub-system"
help = "An error has been detected in the Analog-to-Digital converter
subsystem. Please record the hexadecimal number listed with the error and
cycle power. If the error persists, contact the nearest Agilent Technologies
Instrument support center for assistance.
desc = "Error detected in slit positioning system"
help = "An error has been detected in the motor which controls the resolution
bandwidth slit wheel. Please record the hexadecimal number listed with the
error and cycle power. If the error persists, contact the nearest Agilent
Technologies Instrument support center for assistance.
desc = "Error detected in grating positioning system"
help = "An error has been detected in the motor which controls the diffraction
grating. Please record the hexadecimal number listed with the error and cycle
power. If the error persists, contact the nearest Agilent Technologies Instrument
support center for assistance.
desc = "Trans-Impedance calibration data is invalid"
help = "Factory calibration data for the trans-impedance amplifier is invalid.
Please record the hexadecimal number listed with the error and cycle power. If
the error persists, contact the nearest Agilent Technologies Instrument support
center for assistance.
desc = "Sweep Timeout"
help = "A sweep was started but did not finish in the expected amount of time.
The trace data acquired during this sweep may not be valid. Try taking another
sweep. If the error persists, contact the nearest Agilent Technologies Instrument
support center for assistance.