Programming Commands
CALCulate Subsystem Commands
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:DELta[:STATe]
Turns the delta marker function on or off for a particular marker.
Individual markers can have only one marker function on at a time.
Turning the delta function for a marker on will turn any other marker function
off. For example, turning the delta function on for a marker that has the band-
width function on, will turn the bandwidth function off and turn the delta
function on for the marker.
If the delta function is turned on for a marker that is off, the marker will be
turned on, placed at the center wavelength, and the delta function will be
turned on.
Read the offset: CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion
:DELTa:X:OFFSet?
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n : D E LTa
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n : D E LTa
Delta Marker On|Off
Related Key
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:DELTa:X:OFFSet?
Returns the difference between the absolute X-axis value of the delta marker
and the X-axis value of the reference marker.
The units of the value returned by the query are determined by the
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:DELTa:X:READout state.
For READout of FREQuency, the units are Hertz. For READout of WAVe-
length, the units are meters. For READout of TIME, the units are seconds.
This query generates a "Settings conflict" error if the delta function is
the specified marker.
Syntax
: C A L C
6-40
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[ : S TATe ] O F F | O N | 0 | 1
[ : S TATe ] ?
ulate[1|2|3|4|5|6]
: M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n : D E LTa : X
: O F F S e t ?
for
OFF