Specifications and Regulatory Information
Wavelength Specifications
Wavelength Specifications
Agilent Technologies warrants that the following specifications will be met
under the following instrument operating conditions:
Unless otherwise indicated,
• the specified temperature range is 10 ° to 30 ° C
• only over the 1510 to 1580 nm wavelength range
• device and fixtures present at least a 30 dB return loss to the laser source
• normalization is repeated when the temperature changes 4 ° C since the last
normalization or at least once every 24 hours and normalization is performed
at the same sweep mode as the measurement
• wavelength calibration is repeated when the temperature changes 4 ° C from
the last calibration or at least once every 24 hours
• laser warm up of 90 minutes prior to performing measurements
• sweep span is less than 50 nm
• Acc Unspec notation not present on display
• specifications assume sufficient sensitivity used to eliminate noise
9-8
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com