CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:MAXimum
Remote
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:SCENter
Commands
SENSe:WAVelength:CENTer
Peak Excursion (Marker Setup Panel)
Sets the peak excursion value for the marker search functions.
For marker search functions, a signal peak is defined as a rise and fall in the
displayed response by at least the peak excursion value. If peak excursion is
set too high, legitimate peaks may not be discerned as signals. With lower val-
ues more signals will be discerned, but peaks in the noise floor may also be
discerned as signals. To reduce the noise floor variance to a value less than the
peak excursion rate, try using video averaging and/or lowering the sensitivity.
Key Path
Markers > Marker Setup > Peak Excursion
Marker Search Threshold, Next Peak Down ↓ , Next Peak Right → , Next
Related Functions
Peak Left ←
Remote
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:PEXCursion:PEAK
Commands
Peak to REF LEVEL
Sets the value of the reference level equal to the value of the highest point on
the active trace.
Amplitude > Peak to REF LEVEL
Key Path