Agilent Technologies 86082A User Manual page 290

Wavelength domain component analyzer
Table of Contents

Advertisement

If the number of measurements is greater than or equal to the count, the fol-
lowing formula is used to calculate the data:
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : AV E R a g e : C O U N t < n u m e r i c _ v a l u e >
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : AV E R a g e : C O U N t ?
Averaging On|Off
Related Key
:CALCulate[1|2|3|4|5|6]:AVERage[:STATe]
Turns trace averaging on and off. When using this command, you must have
the selected trace on. [1|2|3|4|5|6] refers to traces A, B, C, D, E, and F respec-
tively.
If a math expression with the corresponding CALCulate subopcode is off, the
SENSe:DATA is averaged. If the math expression is on, the result of the math
expression is averaged.
The CALCulate[1|2|3|4|5|6]:AVERage, CALCulate[1|2|3|4|5|6]:MAXimum, and
:CALCulate[1|2|3|4|5|6]:MINimum states are mutually exclusive. Each trace
can have only one of these functions on at a time. Turning
:CALCulate[1|2|3|4|5|6]:AVERage on will automatically turn
CALCulate[1|2|3|4|5|6]:MAXimum and CALCulate[1|2|3|4|5|6]:
MINimum off. Each trace can have different CALCulate blocks turned on.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : AV E R a g e [ : S TATe ] O F F | O N | 0 | 1
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : AV E R a g e [ : S TATe ] ?
Averaging On|Off
Related Key
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
count 1
×
------------------------
New average
=
count
Programming Commands
CALCulate Subsystem Commands
new measurement
last average
+
--------------------------------------------------
count
6-29