Status Listings
Analyzer Warnings
Table 7-7. Analyzer Warnings (6 of 12)
Error Number
5056
5057
5058
5059
5060
5061
5062
7-24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Description [description/explanation/examples]
desc = "Trajectory align cannot find input signal"
help = "The trajectory align procedure cannot find a usable input signal. Make
sure you have a signal connected to the optical input."
desc = "Invalid settings for trajectory align"
help = "The external trajectory align function was requested. It cannot be
executed because the start and/or stop wavelength settings are invalid. Start
and stop wavelengths must be between 600nm and 1700nm. In addition the
difference between start and stop wavelengths (span) must be greater than
25nm. To correct this problem adjust the start and/or stop wavelength settings
to be valid."
desc = "Out of memory"
help = "The analyzer has run out of execution memory. An internal function was
aborted due to lack of execution memory. The correction for this problem is to
cycle power."
desc = "Trajectory align: marginal input signal"
help = "The trajectory align procedure was aborted due to an input signal with
marginal amplitude. The input signal was large enough when the function began
it's operation but later became too small. This usually happens when the signal
is on the edge of being too small. To correct this problem increase the input
signal level. If this is not possible then a slight increase in signal level can
sometimes be achieved by cleaning optical fiber connectors."
desc = "Trajectory align failed"
help = "The trajectory align procedure failed. The trajectory adjustments
computed were invalid. This is usually occurs after the analyzer has received a
large mechanical shock. To correct this problem try the trajectory align
procedure again. If the error persists, contact the Agilent support center nearest
your location."
desc = "Invalid marker trace"
help = "The marker could not be positioned. The marker is on a trace which does
not contain any valid data. This is sometimes caused by a trace math result
which is invalid."
desc = "ADC Triggered Sweep Too Fast"
help = "A sweep was taken with one of the ADC trigger modes enabled. The
sweep rate was too fast to allow trace data to be acquired for every
wavelength. To resolve this problem increase sweep time or increase the
frequency the external trigger input signal."