Using Trace Math To Measure Wavelength Drift - Agilent Technologies 86082A User Manual

Wavelength domain component analyzer
Table of Contents

Advertisement

Using Trace Math to Measure Wavelength Drift

This procedure assumes that you have normalized your test setup and posi-
tioned the response on the display. For more information on these processes,
refer to "Normalizing the Measurement" on page 3-11 and refer to "Status
Reporting for Laser Wavelength Zeroing" on page 3-13.
1 Setup your device so that you can view the response on the display.
2 From the front panel, press Traces > Active Trace > TrA.
3 Press Single Sweep to update Trace A then press Update A off.
4 Press Active Trace > TrB.
5 Press Sweep > Repeat Sweep On to continuously update the measured
response on Trace B.
6 Press Traces > Trace Math Off > Default Math Trace C >
Log Math C = A – B.
You can now monitor the wavelength drift of your DUT over time.
Figure 3-17. Top trace shows wavelength drift over time
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com