Status Listings
Analyzer Warnings
Table 7-7. Analyzer Warnings (12 of 12)
Error Number
6753
6754
6755
6756
7998
7999
7-30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Description [description/explanation/examples]
desc = "The window specified is invalid."
help = "The window specified is invalid. The window should be a constant."
desc = "Search failed: no data"
help = "A marker search operation was requested. After clipping the trace data
to screen limits and line marker limits (if enabled) there was no data to search.
This can be caused when all trace data points are beyond current X axis screen
limits or when there are no trace data points between the line markers. Trace
data values which exceed the current Y axis screen limits will not cause this
problem. Adjust the current X axis screen limits and/or move the line markers to
include at least one trace data point."
desc = "Attempt to divide by 0"
help = "At least one point was attempted to be divided by 0. The result for these
divisions has been set to not-a-number. Please realize that further calculations
with not-a-number values are undefined."
desc = "Syntax error; bad token: "
help = "The grammar expression entered cannot be parsed. Please check the
string entered. The bad token attempts to indicate where the error occurred. The
end of line indicates that the analyzer expected more information. Please
consult the manual for additional help."
desc = "Unknown error detected"
help = "An unlisted error was reported by the instrument software. If this error
persists contact Agilent Technologies for assistance."
desc = "The warning list has overflowed"
help = "The warning list has overflowed. The last entries received have been
deleted."