Measuring Insertion Loss And Return Loss Simultaneously - Agilent Technologies 86082A User Manual

Wavelength domain component analyzer
Table of Contents

Advertisement

Using the Analyzer
Measurement Techniques
Measuring Insertion Loss and Return Loss
Simultaneously
By using a coupler you can display both the insertion loss and return loss of a
filter.
The first step is to normalize the return loss path. The DUT will be replaced
with an Agilent 81000BR reference reflector. The 81000BR is a gold-plated
HMS-10 connector for use in measuring return loss of optical connectors. It
allows you to establish a precise reference for reflection measurements.
Return loss is specified at 0.18 dB +/-0.1 dB (96% +/-2%).
1 Connect the equipment as shown in Figure 3-9.
Figure 3-9. Normalize Receiver Input 2 setup
2 Press Normalize Ref to access the Normalization menu.
3 Press Normalize Setup and make changes as applicable for your device.
4 Press Normalize Input 2 to normalize the return loss path for the device.
Connecting the coupler output to the source output is similar to placing a
mirror at the coupler input (100% reflections), therefore we have a reflection
of a known value.