:CALibration:WAVelength:DATE?
Returns the date of the most recent wavelength calibration.
Syntax
: C A L i b r a t i o n : WAVe l e n g t h : D AT E ?
:CALibration:WAVelength:INTernal
Performs the wavelength calibration using the internal calibrator.
Note
Output 2 on the TLS front panel must be connected to the Source Input before sending
this command.
Syntax
: C A L i b r a t i o n : WAVe l e n g t h : I N Te r n a l
Related Key
Perform Calibration
:CALibration:WAVelength:STATe
Specifies whether or not the calibration wavelength data is applied. Wave-
length accuracy is only specified with wavelength calibration on. Performs
wavelength calibration and TLS settling routine to ensure absolute wavelength
accuracy.
Syntax
: C A L i b r a t e : WAVe l e n g t h : S TATe O N | O F F | 1 | 0
: C A L i b r a t e : WAVe l e n g t h : S TATe ?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Programming Commands
CALibration Subsystem Commands
6-67