New GPIB Address (Remote Setup Panel)
Allows you to enter a new GPIB address. To change the address, use the
numeric entry keys or knob. Press the Defaults softkey to reset the address to
the factory preset default.
Key Path
System > More System Functions > Remote Setup
Current GPIB Address
Related Functions
Next Peak Down
Places the active marker on the next highest peak down from the current
marker amplitude. This peak must meet the peak excursion and threshold cri-
teria. If the specified marker is off, it will be turned on and placed at the cen-
ter wavelength. The search for the next maximum will reference that point.
Markers > More Marker Functions > Marker Search Menu > Next Peak ↓
Key Path
Remote
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:MAXimum:NEXT
Commands
Next Peak Left
Places the marker on the next peak located at a lower X-axis value (usually
wavelength) than the current marker position. This peak must meet the peak
excursion and threshold criteria. If the specified marker is off, it will be turned
on and placed at the center wavelength or frequency.
Markers > More Marker Functions > Marker Search Menu >Next Peak ←
Key Path
Remote
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:MAXimum:LEFT
Commands