Table 9-1. Wavelength Specifications
Wavelength Range
C Band
a
L Band
b
Absolute Wavelength Accuracy
a
C Band
b
L Band
Relative Accuracy
1 nm delta
20 nm delta
>20 nm delta
Wavelength Resolution
(characteristic)
Wavelength Reproducibility
(characteristic)
a. Option 111
b. Option 112
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Precision Sweep
Normal Sweep
Mode
1515 nm to 1575 nm
1480 nm to 1575 nm (typical)
1520 nm to 1620 nm
±7 pm
±7 pm
±5 pm (typical)
±5.1 pm (typical)
±
±
9 pm
9 pm
±
±
7 pm (typical)
7 pm (typical)
±2 pm
± 2 pm
±4 pm
± 4 pm
absolute wavelength
absolute wavelength
accuracy applies
accuracy applies
± 1.0 pm
± 1.0 pm
Specifications and Regulatory Information
Wavelength Specifications
Fast Sweep
Mode
Mode
±9 pm
±7 pm (typical)
±
11.1 pm
±
9.1 pm (typical)
±3 pm
±5 pm
absolute wavelength
accuracy applies
9-9