Saving Measurement and Trace Data

1 Press the front-panel Save/Recall key.
2 Press the Save Menu.... softkey.
3 The Save Setup panel opens. Refer to "To Fill In a Setup Panel" on page 3-2 for
information on changing and selecting items in the setup panel.
The Save Setup panel
Setup panel
Save
selections
Saves the measurement data in a binary format (.dat file). This includes the
traces and all measurement conditions. The .dat file format can only be read
by the WDCA. You will not be able to view this file on your PC. When the file is
recalled, the instrument state will be set to the same state as when the file was
saved.
The Trace(s) Only files are saved in comma separated variable (.csv) format
and are auto named starting with TR_00000.csv. State files are auto named
starting with ST_00000.csv.
When the file is recalled, the trace data will be displayed under the current
instrument settings.To view the instrument settings, press System > More
System Functions > State Info.
Save Traces
Selects the traces to be saved.