Programming Commands
CALCulate Subsystem Commands
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:SRANge:UPPer?
Returns the upper limit for the marker search range.
The range used for the marker search range is the same range used for the
trace mean range. The return value is in meters, unless span is set to zero, in
which case the return value is in seconds.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M a r k e r [ 1 | 2 | 3 | 4 ] : S R A N g e : U P P e r ?
Line Markers Off
Related Key
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:SRANge:UPPer:FREQuency
Sets the upper limit for the marker search range.
Setting this value when :CALCulate[1|2|3|4|5|6]:Marker[1|2|3|4]:SRANge:STATe
is
OFF
SRANge:STATe on. The range used for the marker search range is the same
range used for the trace mean range. Sending the command when the instru-
ment is in a zero span will generate a "Settings conflict" error. The default
units for the parameter is in Hertz.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M a r k e r [ 1 | 2 | 3 | 4 ] : S R A N g e : U P P e r
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:SRANge:UPPer:TIME
Sets the upper limit for the marker search range.
Setting this value when :CALCulate[1|2|3|4|5|6]:Marker[1|2|3|4]:SRANge:STATe
is off will automatically turn :CALCulate[1|2|3|4|5|6]:Marker[1|2|3|4]:
SRANge:STATe on. The range used for the marker search range is the same
range used for the trace mean range. Sending this command while span is not
set to zero will result in a "Settings conflict" error. Default units for the param-
eter is in seconds.