Table 7-7. Analyzer Warnings (7 of 12)
Error Number
5063
5064
5065
5066
5067
5068
5069
5070
5071
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Description [description/explanation/examples]
desc = "Key disabled during applications"
help = "The key you pressed is not active while an application is running. Exiting
the application should re-enable the key."
desc = "Invalid measurement file"
help = "An attempt was made to recall a measurement file which is contains
invalid or corrupt data."
desc = "Error detected in DSP sub-system"
help = "The Digital Signal Processor has reported an unexpected error. Please
record the hexadecimal number listed with the error and cycle power. If the error
persists, contact the Agilent support center nearest your location."
desc = "Error in Enhanced Wavelength Calibration"
help = "Enhanced Wavelength Calibration failed. Check OSA system. IF the error
persists, contact the Agilent support center nearest your location."
desc = "OSNR marker not allowed"
help = "A request was made to enable the OSNR marker readout. The readout
cannot be enabled for one of the following reasons. OSNR markers cannot be
used on traces which have X axis units other than meters. This typically occurs
in zero span where the X axis units are seconds. OSNR markers cannot be used
on traces which have Y axis units other than power. This typically happens when
trace math is non and the trace Y axis represents a ratio."
desc = "Wrong trace X axis units for pit to center"
help = "The pit-to-center function was requested. This function will only execute
if the active trace X axis has units of meters (
trace to one with an X axis in meters or re-measure the active trace in a non-
zero span."
desc = "No application found - Key disabled"
help = "Currently no application loaded in the system."
desc = "Invalid Start & Stop wavelengths"
help = "Invalid start and stop wavelengths. Please check start and stop
wavelengths in Normalization setup screen.
desc = "Invalid Normalization Step Size"
help = "Normalization step size greater than span. Please check start, stop, and
step size wavelengths in Normalization setup screen."
Status Listings
Analyzer Warnings
µ
m, nm, pm). Change the active
7-25