Viewing Multiple Messages

Although multiple errors may exist, the spectrum analyzer displays
only one error message at a time. To view any additional messages, do
the following:
1. Press
and
RECALL
2. Press
RECALL ERRORS
function block.
3. Use the up and down step keys to scroll through any other error
messages which might exist, making note of each error code.

Error Message Elimination

When an error message is displayed, always perform the following
procedure:
1. Press
and
SAVE
SAVE STATE
2. Store the current state in a convenient STATE register. (It may be
necessary to set
SAVELOCK
3. Press
and
CAL
REALIGN LO &IF
4. Press
and
RECALL
5. Recall the previously stored STATE.
6. If an error message is still displayed, refer to the following list of
error messages for an explanation of the error.

System Analyzer Programming Errors (100 to 150)

Refer to the 8560 E-Series Spectrum Analyzers User's Guide for
information on programming the spectrum analyzer.
100 NO PWRON Power-on state is invalid; default state is loaded. Press
SAVE
101 NO STATE State to be RECALLed not valid or not SAVEd.
102 # ARGMTS Command does not have enough arguments.
103 # ARGMTS Command does not have enough arguments.
104 # ARGMTS Command does not have enough arguments.
105 # ARGMTS Command does not have enough arguments.
106 ABORTED! Current operation is aborted; GPIB parser reset.
107 HELLO ?? No GPIB listener is present.
108 TIME OUT Spectrum analyzer timed out when acting as
controller.
Chapter 7
.
MORE 1 OF 2
. An error message is displayed in the active
.
to OFF.)
. Wait for the sequence to finish.
.
RECALL STATE
,
to clear error message.
PWR ON STATE
General Troubleshooting
Error Messages
327