Programming Commands
CALCulate Subsystem Commands
Whether or not interpolation is turned on or off.
Query Response
(Marker Setup) Bandwidth Interpolation On|Off
Related Key
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:BWIDth
|BANDwidth:NDB
Sets the desired vertical offset from the numbered marker of the bandwidth
markers. Based on the type, sets the number of dB (NDB) for passband band-
width marker or notch bandwidth marker. The parameter units are as speci-
fied in the UNIT:RATio command.
This value can be set or queried anytime. The marker does not have to be on
or in the bandwidth function.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n : B W I D t h
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n : B W I D t h
Related
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n : B W I D t h
Commands
(Marker BW On Off) -0.5 dB | -3 dB | -6 dB | -10 dB | -20 dB
Related Key
(Notch Marker On Off) -0.5 dB | -3 dB | -6 dB | -10 dB | -20 dB
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:BWIDth
|BANDwidth: READout
Sets the X-axis readout for frequency or wavelength when the instrument is in
a non-zero span.
This setting controls only the bandwidth marker X-axis readouts and the
X:Left? and X:Right? queries. The delta markers have their own setting. This
setting controls all four bandwidth markers.
6-34
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
| B A N D w i d t h : N D B < n u m e r i c _ v a l u e > F i v e C a s e s : U P,
D O W N , M A X i m u m , M I N i m u m , D E F i n e , VA L u e
| B A N D w i d t h : N D B ?
: T Y P E PA S S / N O T C h