Opens a window displaying warnings. You can print the queue, clear the
queue, or page up and down to view the information. If no warnings are gener-
ated, the function will be grayed out.
Key Path
System > Help > Show Warnings
Show Hardware Errors, Show Critical Errors, Show Notices, or Revision
Related Functions
Remote
SYSTem:ERRor?
Commands
Single Sweep
Initiates one sweep of the measurement range. Use this function to update the
displayed measurement data.
Trigger conditions must be met in order for a single sweep to occur. The
SWEEP indicator light on the front panel is on when the sweep is in progress.
The indicator is off when the sweep is complete.
Tip: Single sweep mode is especially useful when programming the instru-
ment and under the following circumstances:
• insuring the trace reflects current measurement range settings
• capturing traces before processing them with math commands
• capturing traces before positioning markers
Key Path
Sweep > Single Sweep
Single Sweep
Repeat Sweep
Related Functions
Remote
INITiate:IMMediate
Commands