Agilent Technologies 86082A User Manual page 197

Wavelength domain component analyzer
Table of Contents

Advertisement

Remote Operation
Monitoring the Instrument
*SRE common command
Standard Status Structure
The Standard Status Structure monitors the following instrument status
events: operation complete, query error, device dependent error, execution
error, and command error. When one of these events occurs, the event sets
the corresponding bit in the register.
Command
*ESR? common command
*OPC common command
*ESE common command
Operation Status Structure
Contains bits that report on the instrument's normal operation. Additional
transition filters give you the ability to select the logic transitions which set
the event register bits. For example, you can define the Measuring bit of the
Operation Status Structure to be set when a status line transitions from false
to true. This is a positive transition. You can also specify a negative transition
where bits are set when a status line transitions from true to false.
Command
STAT:OPER:COND?
STAT:OPER:EVEN?
STAT:OPER:ENAB
STAT:OPER:NTR
STAT:OPER:PTR
5-22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Sets or reads the event enable register value
(mask).
Use
Returns and clears the value of the event regis-
ter.
When all operations have finished, sets bit 0 of
the event register. The query returns a 1 when
all operations have finished.
Sets or returns the value of the event enable reg-
ister (mask).
Use
Returns the value of the condition register.
Returns the value of the event register.
Sets or returns the value of the event enable reg-
ister (mask).
Sets or returns the value of the negative transi-
tion register.
Sets or returns the value of the positive transi-
tion register.