The point does not have to meet the pit excursion and threshold criteria. The
marker trace is determined by the
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:TRACe command. If the specified
marker is off, it will be turned on and placed on the lowest point of the trace.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m
Related
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m : L E F T
Commands
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m : N E X T
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m : R I G H t
Pit Search
Related Key
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:MINimum:LEFT
Places the marker on the next pit located at a shorter wavelength than the
current marker wavelength position.
This next pit must meet the pit excursion and threshold criteria. If the speci-
fied marker is off, it will be turned on, placed at the center wavelength, and
the search to the left will begin from that point.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m : L E F T
Related
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m : N E X T
Commands
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : M I N i m u m : R I G H t
Next Pit Left ←
Related Key
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:MINimum:NEXT
Places the marker on the next lowest pit from the current marker amplitude.
This next lowest pit must meet the pit excursion and threshold criteria. If the
specified marker is off, it will be turned on, placed at the center wavelength,
and the search for the next minimum will begin from that point.