Normalize Start Wavelength (Normalize Setup Panel)
Sets the starting point of the normalized wavelength measurement range. The
normalization takes multiple sweeps on linked together traces to make a nor-
malization trace.
Tip: Make sure that you choose a normalized wavelength range that is equal to
or greater than the measurement wavelength range.
Normalize Reference > Normalize Setup > Start Wavelength
Key Path
Related Functions
Normalize Input 1, Normalize Setup
Remote
SENSe:NORMalize:WAVelength:STARt
Commands
Normalize Stop Wavelength (Normalize Setup Panel)
Sets the stopping point of the normalized wavelength measurement range.
The normalization takes multiple sweeps on linked together traces to make a
normalization trace.
Tip: Make sure that you choose a normalized wavelength range that is equal to
or greater than the measurement wavelength range.
Key Path
Normalize Reference > Normalize Setup > Stop Wavelength
Related Functions
Normalize Input 1, Normalize Setup
SENSe:NORMalize:WAVelength:STOP
Remote
Commands