Table 7-7. Analyzer Warnings (11 of 12)
Error Number
6742
6744
6745
6746
6747
6748
6749
6750
6751
6752
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Description [description/explanation/examples]
desc = "Requested amplitude not found"
help = "A math expression to search for a specific amplitude in a trace did not
succeed. There are no trace points with the desired amplitude."
desc = "Excursion should be in dB"
help = "The units for the excursion are not in dB."
desc = "Log of a negative number is not allowed."
help = "During evaluation of a math expression the logarithm of a negative
number was encountered."
desc = "Math expression contains a circular reference."
help = "A new math expression was entered. The expression was rejected
because it would create a circular reference. For example if trace math for trace
C is set to (A-B) then setting trace math for trace B to (C+D) would create a
circular reference."
desc = "Next peak not found"
help = "A next-peak search was requested. There are no more peaks on the
current marker's trace in the requested direction. To find additional peaks
reduce the marker peak excursion setting or adjust sensitivity."
desc = "Next pit not found"
help = "A next-pit search was requested. There are no more pits on the current
marker's trace in the requested direction. To find additional pits reduce the
marker pit excursion setting or adjust sensitivity."
desc = "Peak not found"
help = "A peak search was requested. There are no valid peaks on the current
marker's trace. Try decreasing the marker peak excursion setting or adjust the
sensitivity setting."
desc = "Pit not found"
help = "A pit search was requested. There are no valid pits on the current
marker's trace. Try decreasing the marker pit excursion setting or adjust the
sensitivity setting."
desc = "Trace has no centroid"
help = "The trace has no centroid. This may occur because the sum of trace
points' amplitudes is zero or there a no trace points."
desc = "The reference point is outside trace bounds"
help = "The reference point's wavelength is either too small or too large. The
reference point will be clipped to a trace endpoint."