Normalize Input 1
Initiates and sets the normalization trace on input1. Normalization is used to
observe changes to a displayed response. For example, you might want to nor-
malize your test setup prior to measuring the device. When the device is mea-
sured the response will be corrected for any losses due to the test setup.
Normalization should be performed after each 8 hours of operation.
Tip: If you accidently start the normalization before connecting a patch cord
between the TLS output and receiver input, you can abort by pressing the
Preset key. The instrument will take approximately 10 seconds to respond.
N o t e
Normalization data is valid only for the sweep mode used during normalization.
For example, if Normal Sweep Mode is the active sweep mode during
normalization, the normalized data will be valid only for normal sweep mode.
Refer to "Normalizing the Measurement" on page 3-11.
Key Path
Normalize Reference > Normalize Input 1
Related Functions
Normalize Input 2, Normalize Setup
Remote
INITiate:NORMalize
Commands
Normalize Input 2
Initiates and sets the normalization trace on input 2. Normalization is used to
observe changes to a displayed response. For example, you might want to nor-
malize your test setup prior to measuring the device. When the device is mea-
sured the response will be corrected for any losses due to the test setup.
Normalization should be performed after each 8 hours of operation.
Tip: If you accidently start the normalization before connecting a patch cord
between the TLS output and receiver input, you can abort by pressing the
Preset key. The instrument will take approximately 10 seconds to respond.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Function Reference
Normalize Input 1
2-35