:INITiate[:IMMediate]
Takes a sweep with the current WDCA settings and displays a trace.
Syntax
: I N I T i a t e [ : I M M e d i a t e ]
Related
: I N I T i a t e : C O N T i n u o u s O F F | O N | 0 | 1
Commands
Single Sweep
Related Key
:INITiate:NORMalize
Initiates a normalization trace on the selected input.
After a through cable is connected, use this command to perform a normaliza-
tion of the selected input.
N o t e
Normalization data is valid only for the sweep mode used during normalization.
For example, if Normal Sweep Mode is the active sweep mode during
normalization, the normalized data will be valid only for normal sweep mode.
Syntax
: I N I T i a t e : N O R M a l i z e 1 | 2
Related
[ : S E N S e ] : N O R M a l i z e : WAVe l e n g t h : S TA R t [ N M | P M ]
[ : S E N S e ] : N O R M a l i z e : WAVe l e n g t h : S T O P [ N M | P M ]
Commands
[ : S E N S e ] : N O R M a l i z e : E N A B l e 1 | 2 , O N | O F F | 0 | 1
Normalize Input 1
Related Key
Normalize Input 2