:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:X:FREQuency
S
ets the X-axis value of the normal marker.
When the delta function is on, the absolute X-axis value of the delta marker is
controlled. When the bandwidth function is on, the X-axis value of the center
marker is controlled.
Sending the command when the specified marker is off will turn the marker on
and place the marker at the desired position. Sending the command when the
instrument is in a zero span will generate a "Settings conflict" error.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : X : F R E Q u e n c y
< n u m e r i c _ v a l u e >
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:X:READout
Sets the X-axis readout for frequency or wavelength when the instrument is in
a non-zero span.
This setting controls only the normal marker X-axis and the delta reference
readout. The bandwidth and delta offset markers have their own settings. This
setting controls all four normal markers.
Trying to set the READout to TIME when in a non-zero span generates a "Set-
tings conflict" error. Trying to set the READout to FREQuency or WAVelength
when in zero span also generate a "Settings conflict" error. When the instru-
ment is set to zero span, the readout will automatically change to TIME. This
command is primarily useful for non-zero spans.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M a r k e r [ 1 | 2 | 3 | 4 ] : X : R E A D o u t
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M a r k e r [ 1 | 2 | 3 | 4 ] : X : R E A D o u t ?
Normal Marker Units
Related Key
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
: F R E Q u e n c y | WAVe l e n g t h | T I M E
Programming Commands
CALCulate Subsystem Commands
6-55