Getting Started
Setting Up the WDCA
N O T E
All specifications apply after the instrument's temperature has been stabilized
after 1 hour of continuous operation.
4 For Option 201 only, upgrade your TLS firmware using the firmware upgrade
procedure provided on the 86082A CD-ROM.
5 Perform a wavelength calibration by pressing System > Wavelength Calibration
> Perform Calibration. The calibration takes approximately 5 minutes.
6 Normalize Receiver Input 1 by pressing Norm Ref > Normalize Input 1.
7 When the normalization is complete, the software automatically takes a sweep.
8 Verify that the normalized trace is at 0 dB. You are now ready to start making
measurements. Continue with "Using the Analyzer" on page 3-1.
1-36