Function Reference
Normalize Reference
N o t e
Normalization data is valid only for the sweep mode used during normalization.
For example, if Normal Sweep Mode is the active sweep mode during
normalization, the normalized data will be valid only for normal sweep mode.
Refer to "Normalizing the Measurement" on page 3-11.
Key Path
Normalize Reference > Normalize Input 2
Related Functions
Normalize Input 1, Normalize Setup
Remote
INITiate:NORMalize
Commands
Normalize Reference
Accesses the following trace normalization functions:
• Normalize Setup
• Normalize Input 1
• Normalize Input 2
• Apply Input 1 Normalization
• Apply Input 2 Normalization
Normalize Setup
Accesses the Normalization Receiver Input Setup panel used to specify the
normalized wavelength range.
Key Path
Normalize Reference > Normalize Setup
Normalize Input 1, Normalize Setup
Related Functions
Remote
Start wavelength: SENSe:NORMalize:WAVelength:STARt
Commands
Stop wavelength: SENSe:NORMalize:WAVelength:STOP
2-36