:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:DELTa:X:OFFSet:
FREQuency
Allows you to set the marker offset in frequency units.
The marker X-axis value corresponds to the reference X value + the offset
value. The default units of the parameter for this command are Hertz.
This query generates a "Settings conflict" error if the delta function is off for
the specified marker.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n
:CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:DELTa:X:OFFSet:
TIME
Allows you to set the marker offset when the instrument is in zero span.
The marker X-axis value corresponds to the reference X value + the offset
value. The default units of the parameter are seconds.
This query generates a "Settings conflict" error if the delta function is off for
the specified marker.
Note: This command functions only in zero span. Use the following steps:
1 Put the instrument in 0 span.
2 Turn on the delta marker with the command:
CALCulate[1|2|3|4|5|6]:MARKer1:FUNCtion:DELTa:STATe ON.
3 Set the time value:
CALCulate[1|2|3|4|5|6]:MARKer1:FUNCtion:DELTa:X:OFFSet:TIME 30 ms
4 Read the offset:
CALCulate[1|2|3|4|5|6]:MARKer1:FUNCtion:DELTa:X:OFFSet?
Syntax
: C A L C u l
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
: D E LTa : X : O F F S e t : F R E Q u e n c y < n u m e r i c _ v a l u e >
: D E LTa : X : O F F S e t ?
ate[1|2|3|4|5|6]:
M A R K e r [ 1 | 2 | 3 | 4 ] : F U N C t i o n : D E LTa : X
: O F F S e t : T I M E < n u m e r i c _ v a l u e >
Programming Commands
CALCulate Subsystem Commands
6-41

Advertisement

Table of Contents
loading

Table of Contents