Analyzer Notices - Agilent Technologies 86082A User Manual

Wavelength domain component analyzer
Table of Contents

Advertisement

Status Listings

Analyzer Notices

Analyzer Notices

System control-related error messages or warnings
The analyzer system changed a setting and generated a warning that the oper-
ation was performed.
Table 7-6. System Control Errors or Warnings
Error Number
1000
1001
1002
1003
1004
2999
7-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Description [description/explanation/examples]
desc = "Sensitivity forced to Auto"
help = "Sensitivity has been forced to the Auto setting because another
instrument setting was made which does not permit sensitivity to be in the
Manual setting."
desc = "Old Mezzanine board in system"
help = "This analyzer has an old mezzanine board installed. It is not necessary to
upgrade the mezzanine board but the software will not be able to correctly
determine which light source options are available."
desc = "No Delta Marker Amplitude for"
help = "The delta marker amplitude display has been turned off due to a units
mismatch. The amplitude units of the reference marker are different than those
of the delta marker. Values in differing units cannot be compared."
desc = "Trace length increased"
help = "Trace length has been increased because the wavelength limit function
was turned off. The minimum trace length is larger when the wavelength limit
function is off and trace length was below the new minimum value."
desc = "Incomplete measurement file"
help = "An attempt was made to recall an incomplete measurement file. This
may be caused by recalling a measurement file that was saved using an older
version of firmware. All compatible measurement information was recalled.
desc = "The Notices list has overflowed"
help = "The Notices list has overflowed. The last entries received have been
deleted."

Advertisement

Table of Contents
loading