Agilent Technologies 86082A User Manual page 319

Wavelength domain component analyzer
Table of Contents

Advertisement

Programming Commands
CALCulate Subsystem Commands
:CALCulate[1|2|3|4|5|6]:MATH:STATe
Determines whether or not math processing is done.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M AT H : S TATe O F F | O N | 0 | 1
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M AT H : S TATe
Related Key
Trace Math Off
:CALCulate[1|2|3|4|5|6]:MAXimum:CLEar
Clears the current maximum hold values for the trace and allows a new maxi-
mum hold to occur.
The trace will be initialized to a very negative dBm value (–300 dBm). If the
specified trace is not in the maximum hold state, this command will have no
effect.
Syntax
: C A L C u l a t e [ 1 | 2 | 3 | 4 | 5 | 6 ] : M A X i m u m : C L E a r
Related Key
Reset Min/Max Hold
:CALCulate[1|2|3|4|5|6]:MAXimum[:STATe]
Turns maximum hold on a trace on or off.
The maximum hold operation compares the current amplitude value of each
point on a trace in the current sweep to the corresponding point detected dur-
ing the previous sweep, then stores the maximum value. The
CALCulate[1|2|3|4|5|6]:AVERage, CALCulate[1|2|3|4|5|6]:MAXimum, and
CALCulate[1|2|3|4|5|6]:MINimum states are mutually exclusive. Each trace can
have only one of these functions on at a time. Turning CALCulate[1|2|3|4|5|6]:
MAXimum on will automatically turn CALCulate[1|2|3|4|5|6]:AVERage and
CALCulate[1|2|3|4|5|6]:MINimum off. Each trace is allowed to have a different
CALCulate block turned on. For example, CALCulate1:MAXimum on,
CALCulate2:MINimum on, CALCulate3:AVERage on will put trace A in maxi-
mum hold, trace B in minimum hold and trace C in trace average mode.
6-58