Exchange Traces; Trace Math Add - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

This command does not allow setting all trace points to the same
NOTE
amplitude value by sending just a single value. If you need to set all
trace points to the same value, you must send all 401 values.
Rawtrace data is available with UINT,16 or INT,32 formatting. It is
unitless, returns uncorrected ADC values, and is the fastest method of
obtaining measurement data.
Example:
Remarks:

Exchange Traces

:TRACe:EXCHange <trace_1>,<trace_2>
Exchanges 2 traces, point by point and leaves both in VIEW mode.
Example:
Front Panel
Access:

Trace Math Add

:TRACe:MATH:ADD
<destination_trace>,<source_trace1>,<source_trace2>
Adds the magnitudes of the two source traces together and places the
result in the destination trace.
Chapter 5
:TRAC:DATA TRACE1,#41604<binary trace
data><LF–EOI>
Commands :MMEM:STOR:TRAC and
:MMEM:LOAD:TRAC are used to transfer trace data to,
or from, the internal hard drive or floppy drive of the
instrument.
There are 401 points in a trace. The trace data format
is determined by :FORMat[:TRACe][:DATA], and the
binary data byte order is determined by
:FORMat:BORDer.
If the parameter to the query is LLINE1 or LLINE2, a
very large positive or negative value is returned at any
point outside the range of limit values. A large positive
number is returned for an upper limit, and a large
negative value for lower limits. There is no SCPI short
form for parameters LLINE1|LLINE2.
Trace_1 choices are: TRACE1|2|3
Trace_2 choices are: TRACE1|2|3
:TRAC:EXCH TRACE3,TRACE2
View/Trace, Operations, 1 <–> 3
View/Trace, Operations, 2 <–> 3
Language Reference