OUTPut subsystem

The output subsystem provides information and settings that control the output of the selected
source.
:OUTPut[1]:<function>:SMODe
This command defines the state of the source when the output is turned off.
Type
Command and query
Usage
:OUTPut[1]:<function>:SMODe <state>
:OUTPut[1]:<function>:SMODe?
<function>
<state>
Details
This command sets the state of the output when the source is off for the selected function.
When the Model 2450 is set to the normal output-off state, the following settings are made when the source is
turned off:
The measurement sense is set to 2-wire
The voltage source is selected and set to 0 V
The current limit is set to 10 percent of the full scale of the present current range
OUTPUT OFF is displayed in the Home page Source area
The Source button on the Home page shows the output that will be sourced when the output is turned
on again
When the high-impedance output-off state is selected and the output is turned off:
The measurement sense is set to 2-wire
The output relay opens, disconnecting the instrument as a load
Opening the relay disconnects external circuitry from the inputs and outputs of the instrument. To prevent
excessive wear on the output relay, do not use this output-off state for tests that turn the output off and on
frequently.
The high-impedance output-off state should be used when the instrument is connected to a power source or
another source-measure instrument. In some cases, it may also be appropriate for devices such as capacitors.
2450-901-01 Rev. B/September 2013
Affected by
Recall settings
Instrument reset
Power cycle
Source configuration list
The function to which this setting applies:
Current: CURRent[:DC]
Voltage: VOLTage[:DC]
See the Details below for specifics regarding each of the following options:
NORMal
HIMPedance
ZERO
GUARd
Section 6: SCPI command reference
Where saved
Save settings
Source configuration list
Default value
NORM
6-31