Section 6: SCPI command reference
:TRIGger:TIMer<n>:STARt:GENerate
This command specifies when timer events are generated.
Type
Command and query
Usage
TRIGger:TIMer<n>:STARt:GENerate <b>
TRIGger:TIMer<n>:STARt:GENerate?
<n>
<b>
Details
When this is set to on, a trigger event is generated immediately when the timer is triggered.
When it is set to off, a trigger event is generated when the timer elapses. This generates the event TIMERN.
Example
TRIG:TIM3:STAR:GEN ON
Also see
None
:TRIGger:TIMer<n>:STARt:OVERrun?
This command indicates if an event was ignored because of the event detector state.
Type
Query only
Usage
:TRIGger:TIMer<n>:STARt:OVERrun?
<n>
Details
This attribute indicates if an event was ignored because the event detector was already in the detected state
when the event occurred.
This is an indication of the state of the event detector built into the timer itself. It does not indicate if an overrun
occurred in any other part of the trigger model or in any other construct that is monitoring the delay completion
event. It also is not an indication of a delay overrun.
Example
TRIG:TIM1:STAR:OVER?
Also see
None
6-170
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
Trigger timer number (1 to 4)
Generate a timer event when the timer delay elapses: OFF or 0
Generate a timer event when the timer starts and when the delay elapses: ON or 1
Affected by
Not applicable
Trigger timer number (1 to 4)
Where saved
Save settings
Set trigger timer 3 to generate an event when the
timer starts and when the timer delay elapses.
Where saved
Not applicable
Checks the overrun status on trigger timer 1.
2450-901-01 Rev. B/September 2013
Default value
0 (OFF)
Default value
Not applicable