Trigger:blender<N>:Overrun - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected events (AND)
before signaling an output event.
Example 1
:DIG:LINE3:MODE TRIG, IN
:DIG:LINE5:MODE TRIG, IN
:TRIG:BLEN1:MODE OR
:TRIG:BLEN1:STIM1 DIG3
:TRIG:BLEN1:STIM2 DIG5
Also see
:TRIGger:BLENder<n>:STIMulus<m>
:TRIGger:BLENder<n>:OVERrun?
This command indicates whether or not an event was ignored because of the event detector state.
Type
Query only
Usage
:TRIGger:BLENder<n>:OVERrun?
<n>
Details
Indicates if an event was ignored because the event detector was already in the detected state when the event
occurred. This is an indication of the state of the event detector that is built into the event blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any other
trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
:TRIG:BLEN1:OVER?
Also see
:TRIGger:BLENder<N>:CLEar
6-126
Model 2450 Interactive SourceMeter® Instrument Reference Manual
(on page 6-127)
Affected by
Not applicable
The blender number (1 or 2)
(on page 6-125)
Set digital I/O lines 3 and 5 as trigger in
lines. Generate a trigger blender 1 event
when a digital I/O trigger happens on line 3
or 5.
Where saved
Not applicable
If an event was ignored, the output is 1.
If an event was not ignored, the output is 0.
2450-901-01 Rev. B/September 2013
Default value
Not applicable

Advertisement