Section 6: SCPI command reference

:TRIGger:LOAD:LOOP:SIMPle

This command loads a predefined trigger model configuration.
Type
Command only
Usage
:TRIGger:LOAD:LOOP:SIMPle <count>
:TRIGger:LOAD:LOOP:SIMPle <count>, <delay>
:TRIGger:LOAD:LOOP:SIMPle <count>, <delay>, <reading buffer>
<count>
<delay>
<readingBuffer>
Details
This command sets up a loop that sets a delay, makes a measurement, and then repeats the loop the number of
times you defined in the count parameter.
Example
*RST
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SOUR:FUNC VOLT
SOUR:DEL 0.1
SOUR:VOLT 5
SOUR:VOLT:ILIM 0.01
TRIG:LOAD:LOOP:SIMP 10
OUTP ON
INIT
*WAI
TRAC:DATA? 1, 10, "defbuffer1", SOUR, READ, REL
Also see
None
6-164
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Not applicable
The number of measurements to make
The time before the measurement (50 µs to 10000 seconds); default is 0
for no delay
A string that indicates the reading buffer; the default buffers (defbuffer1
or defbuffer2) or the name of a user-defined buffer; if no buffer is
specified, this parameter defaults to defbuffer1
Where saved
Not applicable
Reset the instrument and set it to
measure current with automatic range
setting.
Source 5 volts with a source delay of
0.1 s.
Set a current limit of 0.01 A.
Set a simple trigger loop with a count of
10.
Turn the output on.
Start the trigger model.
Postpone execution of subsequent
commands until all previous commands
are finished.
Read data and store the source, reading,
and relative time.
2450-901-01 Rev. B/September 2013
Default value
Not applicable