Digital:line<N>:State - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Section 6: SCPI command reference
:DIGital:LINE<n>:STATe
This command sets a digital I/O line high or low.
Type
Command and query
Usage
:DIGital:LINE<n>:STATe <state>
:DIGital:LINE<n>:STATe?
<n>
<state>
Details
When a reset occurs, the digital line state can be read as high because the digital line is reset to a digital input
(see the mode command). A digital input floats high if nothing is connected to the digital line.
Set the state to zero (0) to clear the bit; any non-zero value sets the bit.
Example
:DIG:LINE1:STAT 1
Also see
Digital I/O port
:DIGital:LINE<n>:MODE
:DIGital:READ?
:DIGital:WRITe <n>
:TRIGger:DIGital<n>:IN:EDGE
6-20
Affected by
Not applicable
Digital I/O trigger line (1 to 6)
Bit low: 0
Bit high: 1
(on page 3-62)
(on page 6-19)
(on page 6-21)
(on page 6-21)
(on page 6-150)
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Where saved
Nonvolatile memory
Sets line 1 (bit B1) of the digital I/O port high.
Default value
See Details
2450-901-01 Rev. B/September 2013