Section 8: TSP command reference
trigger.model.setblock() — trigger.BLOCK_CONFIG_NEXT
This attribute recalls the settings at the next index point of a source or measure configuration list.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.model.setblock(blockNumber, trigger.BLOCK_CONFIG_NEXT, configurationList)
blockNumber
configurationList
Details
When the trigger model reaches a configuration recall next building block, the settings at the next index point in a
configuration list are restored.
Each time this block is encountered, the settings at the next index point in the configuration list are recalled and
take effect before the next step executes. When the last index point in the list is reached, it returns to the first
point.
Example
trigger.model.setblock(5, trigger.BLOCK_CONFIG_NEXT, "measTrigList")
Configure trigger block 5 to load the next index point in the configuration list named measTrigList.
Also see
Configuration lists
8-202
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Restore configuration
Instrument reset
Power cycle
(on page 3-33)
Where saved
Default value
Configuration script Not applicable
The sequence of the block in the trigger model
The configuration list from which to recall settings