Application Note
Se ries

Introduction

To measure the leakage current or insulation resistance of a
device, you need to apply a fixed voltage to the device and
measure the resulting current. Depending on the device under
test, the measured current is typically very small, usually less
than 10nA.
This application consists of two examples that demonstrate:
• How to use the Model 2450 to perform leakage current
measurements on a capacitor
• How to use the Model 2450 to measure insulation resistance
between the two conductors of a coaxial cable
The only difference between these two application examples
is that when you measure leakage current, the results are
returned in units of amperes. When you measure insulation
resistance, the results are returned in units of ohms.
The leakage current application applies the voltage for a
specified period because the device needs time to charge. In
some cases, the resulting current is measured the entire time the
device is biased. In other cases, only one reading is made at the
end of the soak period.
The following topics describe how to do these applications
from the front panel. They also show how to do them using a
remote interface with SCPI commands or Test Script Processor
(TSP
) commands.
®

Equipment required

• One Model 2450 Interactive SourceMeter® instrument
• Two triaxial cables
• One capacitor for the leakage current application
• One coaxial cable or other device for the insulation resistance
application
• One ethernet, GPIB, or USB cable for the TSP and SCPI
remote command examples

Set up remote communications

You can run this application from the front panel or any of the
supported communication interfaces for the instrument (GPIB,
USB, or ethernet).
The following figure shows the rear-panel connection
locations for the remote communication interfaces.
Number 3262
Leakage Current and Insulation Resistance
Measurements Using the Model 2450
SourceMeter
SMU Instrument
®
Figure 1. Model 2450 remote interface connections.

Device connections

Depending on the device under test (DUT), the current
measurement is typically very small, usually <10nA. Measuring
leakage current and insulation resistance involves measuring very
small values. To get more accurate readings, connect the DUT to
the Model 2450 rear panel with low-noise triaxial cables.
Connect the DUT between the FORCE HI and FORCE LO
terminals of the Model 2450.
Figure 2 shows schematic diagrams. One shows measuring
the leakage current of a capacitor. The other shows measuring
the insulator resistance between two conductors of a
coaxial cable.
Figure 2. Connection schematics for the capacitor leakage and insulation
resistance tests.
The following figures show the rear-terminal connections to
the device under test (DUT) for these applications. If capacitor
leakage measurements are noisy, you may need to use the high
capacitance mode or add a low leakage forward-biased diode in
series with the capacitor.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 2450 and is the answer not in the manual?

Questions and answers

Summary of Contents for Keithley 2450

  • Page 1: Equipment Required

    10nA. This application consists of two examples that demonstrate: • How to use the Model 2450 to perform leakage current measurements on a capacitor Figure 1. Model 2450 remote interface connections.
  • Page 2 To set up the application from the front panel: 1. Use the test leads to make the connection from the capacitor Figure 3. Rear-panel connections for the leakage current test. to the rear panel of the Model 2450, as described in Device connections.
  • Page 3 After the code is executed, the measurement results are displayed in the Instrument Console of Test Script Builder. By default, the Model 2450 is configured to use the SCPI From the Instrument Console, you can copy the data into a command set.
  • Page 4 This application sets the Model 2450 to source 20 V and -- Reset the instrument, which also clears the buffer. makes 10 resistance readings with a 100 ms interval time. Once reset() the measurements are made, the output is turned off.
  • Page 5 --Set up the source function. smu.source.func = smu. to make changes to the example TSP code. FUNC_DC_VOLTAGE smu.source.ilimit.level = 0.1 By default, the Model 2450 is configured to use the SCPI smu.source.level = 20 command set. You must select the TSP command set before --Turn on the source output and take readings.
  • Page 6 Specifications are subject to change without notice. All Keithley trademarks and trade names are the property of Keithley Instruments, Inc. All other trademarks and trade names are the property of their respective companies. A Greater Measure of Confidence KEITHLEY INSTRUMENTS, INC.