Model 2450 Interactive SourceMeter® Instrument Reference Manual
If you have FORMat[:DATA] set to REAL or SREAL, you will have fewer options for buffer elements.
If you request one of the buffer elements, you will see the error 1133, "Parameter 4, Syntax error,
expected valid name parameters."
Example
:TRACe:MAKE "voltMeasBuffer", 10000
:SENSe:FUNCtion "VOLTage"
:READ? "voltMeasBuffer", FORM, DATE, READ
Create a buffer named voltMeasBuffer. Make a measurement, store it in the buffer voltMeasBuffer,
and return the formatted readings, data, and reading buffer elements for the last reading stored in
voltMeasBuffer.
Sample output is:
-00.0020 mV,05/16/2014,-2.031637E-06
Also see
:FETCh?
:INITiate[:IMMediate]
CALCulate subsystem
The commands in this subsystem configure and control the math and limit operations.
:CALCulate[1]:<function>:MATH:FORMat
This command specifies which math operation is performed on measurements.
Type
Command and query
Usage
:CALCulate[1]:<function>:MATH:FORMat <name>
:CALCulate[1]:<function>:MATH:FORMat?
<function>
<name>
2450-901-01 Rev. B/September 2013
(on page 6-3)
(on page 6-124)
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The function to which this setting applies:
Current: CURRent[:DC]
•
Resistance: RESistance
•
Voltage: VOLTage[:DC]
•
The name of the math operation:
y = mx+b: MXB
•
Percent: PERCent
•
Reciprocal: RECiprocal
•
Section 6: SCPI command reference
Where saved
Save settings
Measure configuration list
Default value
PERC
6-7