Model 2450 Interactive SourceMeter® Instrument Reference Manual

:TRIGger:LOAD:CONFiguration:LIST

This command loads a predefined trigger model configuration that uses source and measure configuration lists.
Type
Command only
Usage
:TRIGger:LOAD:CONFiguration:LIST <measureConfigList>, <sourceConfigList>
:TRIGger:LOAD:CONFiguration:LIST <measureConfigList>, <sourceConfigList>, <delay>
:TRIGger:LOAD:CONFiguration:LIST <measureConfigList>, <sourceConfigList>, <delay>,
<readingBuffer>
<measureConfigList>
<sourceConfigList>
<delay>
<readingBuffer>
Details
This trigger model template incorporates a source configuration list and measure configuration list.
You must set up the configuration lists before loading the trigger model.
You can also set a delay and reading buffer.
Example
TRIG:LOAD:CONF:LIST "MyMeasList", "src config"
Load the configuration list trigger model, using the measurement configuration list MyMeasList and the
source configuration list src config.
Also see
None
2450-901-01 Rev. B/September 2013
Affected by
Not applicable
The configuration list that contains the measurement settings to be used
for this trigger model
The configuration list that contains the source settings to be used for this
trigger model
The delay time before the measurement (seconds); default is 0
A string that indicates the reading buffer; the default buffers (defbuffer1
or defbuffer2) or the name of a user-defined buffer; if no buffer is
specified, this parameter defaults to defbuffer1
Section 6: SCPI command reference
Where saved
Not applicable
Default value
Not applicable
6-161