Model 2450 Interactive SourceMeter® Instrument Reference Manual
[:SENSe[1]]:<function>:RELative
This command specifies the relative offset value for the measurement.
Type
Command and query
Usage
[:SENSe[1]]:<function>:RELative <n>
[:SENSe[1]]:<function>:RELative?
<function>
<n>
Details
This command specifies the relative offset value that is used for measurements. When relative offset is enabled,
all subsequent measured readings are offset by the value that is set for this command.
Example
CURR:REL .5
CURR:REL:STAT ON
Also see
Relative offset
[:SENSe[1]]:<function>:RELative:ACQuire
[:SENSe[1]]:<function>:RELative:STATe
2450-901-01 Rev. B/September 2013
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The measurement function to which this setting applies:
Current: CURRent[:DC]
•
Resistance: RESistance
•
Voltage: VOLTage[:DC]
•
The relative offset value:
Current: −1 to 1
•
Resistance: –200000000 to 200000000
•
Voltage: −200 to 200
•
(on page 3-1)
(on page 6-57)
Where saved
Save settings
Measure configuration list
Set the relative offset for current
measurements to 0.5. Enable relative offset.
(on page 6-56)
Section 6: SCPI command reference
Default value
0.0
6-55