Section 6: SCPI command reference
When the zero output-off state is selected, when you turn off the output:
•
The measurement sense is changed to 2-wire sense
•
The source function is set to voltage
•
The source voltage is set to 0
Set the range to the presently selected range (turn off autorange)
•
•
Program the voltage DAC to zero, and the current DAC to full scale of the present current range.
When the zero output-off state is selected, you can use the instrument as an ammeter because it is outputting
0 V.
When the guard output-off state is selected and the output is turned off, the following actions occur:
•
The measurement sense is changed to 2-wire sense
•
The current source is selected and set to 0 A
•
The voltage limit is set to 10% full scale of the present voltage range
If you send this command without the <function> parameter, it will set the output-off state for all
functions.
Example
:OUTP:CURR:SMOD HIMP
Also see
Output-off state
:OUTPut[1][:STATe]
6-32
(on page 2-82)
(on page 6-34)
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Sets the output-off state for the current
function so that the instrument opens the
output relay when the output is turned off.