Smu.measure.configlist.delete() - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual

smu.measure.configlist.delete()

This command deletes a measure configuration list.
Type
TSP-Link accessible
Function
Yes
Usage
smu.measure.configlist.delete(listName)
smu.measure.configlist.delete(listName, point)
listName
point
Details
Deletes a configuration list. If the point parameter is not specified, the entire configuration list is deleted. If the
point parameter is specified, only the specified configuration point in the list is deleted.
Example
smu.measure.configlist.delete("myMeasList")
smu.measure.configlist.delete("myMeasList", 2) Delete configuration point 2 from the measure
Also see
Configuration lists
smu.measure.configlist.create()
2450-901-01 Rev. B/September 2013
Affected by
The name of the configuration list
The number that identifies a specific configuration point on the configuration list to
delete
(on page 3-33)
(on page 8-92)
Section 8: TSP command reference
Where saved
Default value
Delete a measure configuration list named
myMeasList.
configuration list named myMeasList.