Interactive Swipe Screens - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual

Interactive swipe screens

The Model 2450 touch-screen display has multiple screens that you can access by gently swiping left
or right on the lower half of the display. The following topics describe each of these screens.
SETTINGS swipe screen
The SETTINGS screen gives you front-panel access to some instrument settings so that you can
change, enable, or disable them quickly. It also shows what the present settings are.
To disable or enable a setting, select the box next to the setting so that it shows an X (disable) or a
check mark (enable).
2450-901-01 Rev. B/September 2013
Figure 5: SETTINGS swipe screen
Setting
Description
Filter
Shows the present type of the filter (repeating average or
moving average).
Math
Shows the present math function (percent, reciprocal, or
y = mx+b).
Rel
Shows the present value of the relative offset function.
Auto Zero
Sets the instrument so that it automatically returns the
source to a zero value before taking another reading.
NPLC
Sets the number of power line cycles (0.01 to 10.00).
Display Digits
Sets the number of digits visible on the display (3 to 6).
Auto
Turns autoranging on or off.
Range arrows
Moves the autorange setting up or down a level. Press
the up arrow to move up a range; press the down arrow
to move down a range.
Section 2: General operation
2-11

Advertisement

Table of Contents
loading

Table of Contents