Keithley 2450 Reference Manual page 174

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual
Using TSP commands:
Once you create a reading buffer using TSP commands, if you use that buffer name for another
buffer or variable, you can no longer access the original buffer.
The following TSP commands are associated with data storage operation:
Command
buffer.clearstats()
buffer.delete()
buffer.make()
buffer.save()
buffer.saveappend()
bufferVar.capacity
bufferVar.clear()
bufferVar.fillmode
bufferVar.logstate
bufferVar.n
Accessing the data in buffers
Using SCPI commands:
To access a buffer, include the buffer name in the respective command. For example, the following
commands return five readings (including the measurement, source value, and relative time) from a
user-defined buffer named testData:
-- Create a buffer named testData to store 100 readings.
TRAC:MAKE "testData", 100
-- Set the instrument to make 5 readings for all measurement requests.
SENS:COUN 5
-- Make the readings and store them in the buffer
TRAC:TRIG "testData"
-- Read the 5 readings including the measurement, source value, and relative time
for each point.
TRAC:DATA? 1, 5, "testData", MEAS, SOUR, REL
2450-901-01 Rev. B/September 2013
Description
This function clears all statistics from the specified buffer. This function does
not clear the readings. See
This function deletes a user-defined reading buffer. See
page 8-9).
This function creates a user-defined reading buffer. You cannot use this
command on the default buffers. See
see
buffer.make()
(on page 8-11).
This function saves data from the specified reading buffer to a USB flash
drive. See
Storing readings in buffers
(on page 8-12).
This function appends data from the reading buffer to a file on the USB flash
drive. See
buffer.saveappend()
This command reads the number of readings a buffer can store. This allows
you to change the number of readings the buffer can store. See
bufferVar.capacity
(on page 8-15).
This function clears all readings and statistics from the specified buffer. See
Clearing buffers
(on page 3-26). Also see
This attribute determines if a reading buffer is filled continuously or is filled
once and stops. See
bufferVar.fillmode
This attribute indicates whether the reading buffer should log informational
events. See
bufferVar.logstate
This attribute contains the number of readings in the specified buffer. See
bufferVar.n
(on page 8-23).
Section 3: Functions and features
buffer.clearstats()
(on page 8-8).
Creating buffers
(on page 3-12). Also
(on page 3-17). Also see
(on page 8-14).
bufferVar.clear()
(on page 8-19).
(on page 8-22).
buffer.delete()
(on
buffer.save()
(on page 8-17).
3-29