Model 2450 Interactive SourceMeter® Instrument Reference Manual
:CALCulate2:<function>:LIMit<Y>:FAIL?
This command queries the results of a limit test.
Type
Query only
Usage
:CALCulate2:<function>:LIMit<Y>:FAIL?
<function>
<Y>
Details
The results of the limit test for limit Y:
NONE: Test passed; the measurement is between the upper and lower limits
HIGH: Test failed; the measurement exceeded the upper limit
LOW: Test failed; the measurement exceeded the lower limit
BOTH: Test failed; the measurement exceeded both limits
These commands query the result of a limit test for the selected measurement function.
The response message indicates if the limit test has passed or how it failed.
Reading the results of a limit test does not clear the fail indication of the test. To clear a failure, send the clear
command.
Note that if you are making a series of measurements and auto clear enabled for a limit, the last measurement
limit dictates the fail indication for the limit. If autoclear is disabled, you can take a series of readings and read
fails to see if any of one of the readings failed.
To use this attribute, you must set the limit state to enable.
Example
CALC2:VOLT:LIM1:LOW 0.25
CALC2:VOLT:LIM1:UPP 2.5
CALC2:VOLT:LIMIT1:STAT ON
CALC2:VOLT:LIMIT1:FAIL?
Also see
:CALCulate2:<function>:LIMit<Y>:STATe
Limit testing and binning
2450-901-01 Rev. B/September 2013
Affected by
Not applicable
The function to which this setting applies:
Current: CURRent[:DC]
Resistance: RESistance
Voltage: VOLTage[:DC]
The test that the result applies to, 1 or 2
(on page 6-17)
(on page 3-68)
Section 6: SCPI command reference
Where saved
Not applicable
Set lower limit 1 for voltage to 0.25 volts.
Set upper limit 1 for voltage to 2.5 volts.
Enable limit 1 testing for voltage.
Check the test results.
Default value
Not applicable
6-15