[:SENSe[1]]:<function>:RELative:STATe
This command enables or disables the relative offset value.
Type
Command and query
Usage
[:SENSe[1]]:<function>:RELative:STATe <b>
[:SENSe[1]]:<function>:RELative:STATe?
<function>
<b>
Details
This attribute enables or disables relative measurements. When relative measurements are enabled, all
subsequent measured readings are offset by the relative offset value calculated when you acquire the relative
offset value.
Each returned measured relative reading is the result of the following calculation:
Display value = Actual measured value - Relative offset value
Example
SENS:FUNC "VOLT"
SENS:VOLT:REL 5
:SENSe:VOLT:REL:STATe ON
Also see
[:SENSe[1]]:<function>:RELative
[:SENSe[1]]:<function>:RELative:ACQuire
2450-901-01 Rev. B/September 2013
Affected by
Recall settings
Instrument reset
Power cycle
Measure configuration list
The measurement function to which this setting applies:
Current: CURRent[:DC]
Resistance: RESistance
Voltage: VOLTage[:DC]
Disable the relative offset: 0 or OFF
Enable the relative offset: 1 or ON
(on page 6-55)
(on page 6-56)
Section 6: SCPI command reference
Where saved
Save settings
Measure configuration list
Set the measurement function to volts with a
relative offset of 5 V and enable the relative
offset function.
Default value
0 (OFF)
6-57