Source[1]:<Function>:Delay:auto - Keithley 2450 Reference Manual

Interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2450 Interactive SourceMeter® Instrument Reference Manual
:SOURce[1]:<function>:DELay:AUTO
This command enables or disables the autodelay that occurs when the source is turned on.
Type
Command and query
Usage
:SOURce[1]:<function>:DELay:AUTO <state>
:SOURce[1]:<function>:DELay:AUTO?
<function>
<state>
Details
When auto delay is turned on, the actual delay that is set depends on the range.
When source autodelay is on, if you set a source delay, the autodelay is turned off.
Example
SOUR:CURR:DEL:AUTO OFF
Also see
:SOURce[1]:<function>:DELay
2450-901-01 Rev. B/September 2013
Affected by
Recall settings
Instrument reset
Power cycle
Source configuration list
The source function to which this setting applies:
Current: CURRent
Voltage: VOLTage
Disable the source auto delay: ON or 0
Enable the source auto delay: OFF or 1
(on page 6-66)
Section 6: SCPI command reference
Where saved
Save settings
Source configuration list
Turn off auto delay when current is being
sourced.
Default value
1 (ON)