:TRIGger:BLOCk:CONFig:RECall

This command recalls the system settings that are stored in a measure or source configuration list.
Type
Command only
Usage
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, <configurationList>
:TRIGger:BLOCk:CONFig:RECall <blockNumber>, <configurationList>, <point>
<blockNumber>
<configurationList>
<point>
Details
When the trigger model reaches a configuration recall building block, the settings in the specified configuration
list are recalled.
You can restore a specific set of configuration settings in the configuration list by defining the index.
Example
SOUR:CONF:LIST:CRE "biasLevel"
SOUR:FUNC VOLT
SENS:FUNC "CURR"
SOUR:VOLT:LEV 5
SOUR:CONF:LIST:STORE "biasLevel"
TRIG:BLOCK:CONF:RECALL 1, "biasLevel", 1
Also see
Configuration lists
:SOURce[1]:CONFiguration:LIST:STORe
6-138
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
The sequence of the block in the trigger model
The measure or source configuration list to recall
The point in the configuration list to recall; defaults to 1 if not selected
(on page 3-33)
(on page 6-65)
Where saved
Save settings
Create a configuration list named
biasLevel. Set the source function to 5
volts and the measure function to current.
Store the configuration list.
Recall the configuration list as block 1 of the
trigger model.
2450-901-01 Rev. B/September 2013
Default value
Not applicable